Difference in Performance of Fecal Immunochemical Tests With the Same Hemoglobin Cutoff Concentration in a Nationwide Colorectal Cancer Screening Program

作者:Chiang Tsung Hsien; Chuang Shu Lin; Chen Sam Li Sheng; Chiu Han Mo; Yen Amy Ming Fang; Chiu Sherry Yueh Hsia; Fann Jean Ching Yuan; Chou Chu Kuang; Lee Yi Chia*; Wu Ming Shiang; Chen Hsiu Hsi
来源:Gastroenterology, 2014, 147(6): 1317-1326.
DOI:10.1053/j.gastro.2014.08.043

摘要

BACKGROUND %26 AIMS: We investigated whether 2 quantitative fecal immunochemical tests (FITs) with the same cutoff concentration of fecal hemoglobin perform equivalently in identifying patients with colorectal cancer (CRC). METHODS: A total of 956,005 Taiwanese subjects, 50 to 69 years old, participated in a nationwide CRC screening program to compare results from 2 FITs; 78% were tested using the OC-Sensor (n = 747,076; Eiken Chemical Co, Tokyo, Japan) and 22% were tested using the HM-Jack (n = 208,929; Kyowa Medex Co Ltd, Tokyo, Japan), from 2004 through 2009. The cutoff concentration for a positive finding was 20 mg hemoglobin/g feces, based on a standardized reporting unit system. The tests were compared using short-term and long-term indicators of performance. RESULTS: The OC-Sensor test detected CRC in 0.21% of patients, with a positive predictive value of 6.8%. The HM-Jack test detected CRC in 0.17% of patients, with a positive predictive value of 5.2%. The rate of interval cancer rate was 30.7/100,000 person-years among subjects receiving the OC-Sensor test and 40.6/100,000 person-years among those receiving the HM-Jack test; there was significant difference in test sensitivity (80% vs 68%, P = .005) that was related to the detectability of proximal CRC. After adjusting for differences in city/county, age, sex, ambient temperature, and colonoscopy quality, significant differences were observed between the tests in the positive predictive value for cancer detection (adjusted relative risk = 1.29; 95% confidence interval, 1.14-1.46) and the rates of interval cancer (0.75; 95% confidence interval, 0.62-0.92). Although each test was estimated to reduce CRC mortality by approximately 10%, no significant difference in mortality was observed when the 2 groups were compared. CONCLUSIONS: Different brands of quantitative FITs, even with the same cutoff hemoglobin concentration, perform differently in mass screening. Population-level data should be gathered to verify the credibility of quantitative laboratory findings.

  • 出版日期2014-12
  • 单位长春大学