摘要

Background. Plethysmographic measurement of haemoglobin concentration (SpHb), pleth variability index (PVT), and perfusion index (PI) with the Radical-7 apparatus is growing in popularity. Previous studies have indicated that SpHb has poor precision, particularly when PI is low. We wanted to study the effects of a sympathetic block on these measurements. Methods. Twenty patients underwent hand surgery under brachial plexus block with one Radical-7 applied to each arm. Measurements were taken up to 20 min after the block had been initiated. Venous blood samples were also drawn from the non-blocked arm. Results. During the last 10 min of the study, SpHb had increased by 8.6%. The PVT decreased by 54%, and PI increased by 188% in the blocked arm (median values). All these changes were statistically significant. In the non-blocked arm, these parameters did not change significantly. Conclusions. Brachial plexus block significantly altered SpHb, PVI, and PI, which indicates that regional nervous control of the arm greatly affects plethysmographic measurements obtained by the Radical-7. After the brachial plexus block, SpHb increased and PVI decreased.

  • 出版日期2015-5