Background: Thermoluminescent detectors, type MTS-6, containing isotope Li-6 (lithium) are sensitive in the range of thermal neutron energy; the Pu-239-Be (plutonium-and-beryllium) source emits neutrons in the energy range from 1 to 11 MeV. These seemingly contradictory elements may be combined by using the paraffin moderator, a determined density of thermal neutrons in the paraffin block and a conversion coefficient neutron flux to kerma, not forgetting the simultaneous registration of the photon radiation inseparable from the companion neutron radiation. The main aim of this work is to present the idea of calibration of thermoluminescent detectors that consist of a Li-6 isotope, by using Pu-239-Be neutron radiation source. Material and Methods: In this work, MTS-6 and MTS-7 thermoluminescent detectors and a plutonium-and-beryllium (Pu-239-Be) neutron source were used. Paraffin wax fills the block, acting as a moderator. The calibration idea was based on the determination of dose equivalent rate based on the average kerma rate calculated taking into account the empirically determined function describing the density of thermal neutron flux in the paraffin block and a conversion coefficient neutron flux to kerma. Results: The calculated value of the thermal neutron flux density was 1817.5 neutrons/cm(2)/s and the average value of kerma rate determined on this basis amounted to 244 mu Gy/h, and the dose equivalent rate 610 mu Sv/h. The calculated value allowed for the assessment of the length of time of exposure of the detectors directly in the paraffin block. Conclusions: The calibration coefficient for the used batch of detectors is (6.80 +/- 0.42)x10(-7) Sv/impulse.