摘要

A novel on-line preconcentration and determination system based on a fiber-packed column was developed for speciation analysis of Cr in drinking water samples prior to its determination by flame atomic absorption spectrometry (FAAS). All variables involved in the development of the preconcentration method including, pH, eluent type, sample and eluent flow rates, interfering effects, etc., were studied in order to achieve the best analytical performance. A preconcentration factor of 32 was obtained for Cr(VI) and Cr(III). The levels of Cr(III) species were calculated by difference of total Cr and Cr(VI) levels. Total Cr was determined after oxidation of Cr(III) to Cr(VI) with hydrogen peroxide. The calibration graph was linear with a correlation coefficient of 0.999 at levels near the detection limit and up to at least 50 mu g L-1. The relative standard deviation (R.S.D.) was 4.3%(C=5 mu gL(-1) Cr(VI), n = 10. sample volume = 25 mL). The limit of detection (LOD) for both Cr(III) and Cr(VI) species was 0.3 mu g L-1. Verification of the accuracy was carried out by the analysis of a standard reference material (NIST SRM 1643e "Trace elements in natural water"). The method was successfully applied to the determination of Cr(III) and Cr(VI) species in drinking water samples.

  • 出版日期2009-2-15