Assessing agreement between point of care and laboratory results for lipid testing from a clinical perspective

作者:Gialamas Angela*; Laurence Caroline O; Yelland Lisa N; Tideman Philip; Worley Paul; Shephard Mark D; Tirimacco Rosy; Willson Kristyn J; Ryan Philip; Gill Janice; Thomas David W; Beilby Justin J
来源:Clinical Biochemistry, 2010, 43(4-5): 515-518.
DOI:10.1016/j.clinbiochem.2009.11.014

摘要

Objectives: Investigate agreement between lipid pathology results from point-of-care testing (PoCT) devices and laboratories.
Design and methods: Agreement was assessed using the Bland-Altman method.
Results: Mean difference (limits of agreement) were: -0.28 mmol/L (-1.04, 0.48) for total cholesterol, -0.09 mmol/L, (-0.55, 0.36) for HDL-C. Median difference (nonparametric limits of agreement) were 0.07 mmol/L, (-0.40, 3.04) for triglycerides.
Conclusions: The clinical acceptability of the variation between lipid PoCT and laboratory test results is debatable but our work provides baseline data for further research.

  • 出版日期2010-3