A comparison of different recombination methods in mixed radiation fields at high energy accelerators

作者:Zielczyniski M*; Golnik N; Gryzinski M A
来源:Radiation Protection Dosimetry, 2007, 126(1-4): 248-252.
DOI:10.1093/rpd/ncm051

摘要

Recombination chambers and different recombination methods have been used for dosimetry of mixed radiation fields at high-energy accelerators for over 40 years. This paper gives a short overview of 11 selected recombination methods used for the determination of H*(10) in mixed radiation fields at high-energy accelerators. A new correction factor is proposed, mainly in order to take into account the dependence of the chamber sensitivity on radiation quality. This factor depends only on the measurable index of radiation quality and can be determined for a particular chamber during the calibration in a reference field of neutron radiation. A comparison of the results obtained at high-energy accelerators showed that all the methods gave the same values of H*(10), within a specified accuracy of about 20%, so all of them are suitable for monitoring complex mixed radiation fields at workplaces.