Development of Electrolyte-Free Ozone Sensors Using Boron-Doped Diamond Electrodes

作者:Ishii Yuya; Ivandini Tribidasari A; Murata Kazutaka; Einaga Yasuaki*
来源:Analytical Chemistry, 2013, 85(9): 4284-4288.
DOI:10.1021/ac400043b

摘要

The electrochemical detection of dissolved ozone in water was examined using boron-doped diamond (BDD) electrodes. A well-defined reduction peak was observed at similar to 380 mV for H-terminated BDD, whereas it was observed at similar to 200 mV in the case of O-terminated BDD for an ozone solution in a 0.1 M phosphate buffer solution at pH 7. The peak potential for ozone reduction was selective with respect to oxygen reduction at both H- and O-terminated BDD electrodes, whereas it occurred at approximately the same potential as oxygen reduction at other types of solid electrodes, including glassy-carbon, platinum, and gold electrodes. Interference from chlorine was not observed in lower concentration than 300 mu M ClO-. Furthermore, in order to apply the detection technique to electrolyte-free media, BDD microelectrodes were also used. A linear calibration curve for dissolved ozone in water could be achieved between concentrations of 0.49 and 740 mu M, with an estimated detection limit (S/N = 3) of 0.185 mu M (S/N = 3). Excellent stability was demonstrated for repetitions of these calibration curves performed in 3 consecutive days.

  • 出版日期2013-5-7

全文