摘要

Objectives The variable-flow flow driver (FD; EME) and continuous-flow bubble (Fisher-Paykel) continuous positive airway pressure (CPAP) systems are widely used. As these differ in cost and technical requirements, determining comparative efficacy is important particularly where resources are limited. Design We performed a randomised, controlled, equivalence trial of CPAP systems. We specified the margin of equivalence as 2days. We analysed binary variables by logistical regression adjusted for gestation, and log transformed continuous variables by multiple linear regression adjusted for gestation, sex and antenatal steroids. Setting A neonatal unit with no blood gas analyser or surfactant availability and limited X-ray and laboratory facilities Patients Neonates <37weeks of gestation. Interventions We provided CPAP at delivery followed by randomisation to FD or bubble (B). Outcomes Primary outcomeincluded total days receiving CPAP; secondary outcomes included days receiving CPAP, supplemental oxygen, ventilation, death, pneumothorax and nasal excoriation. Results We randomised 125 infants (B 66, FD 59). Differences in infant outcomes on B and FD were not statistically significant. The median (range) for CPAP days for survivors was B 0.8 (0.04 to 17.5), FD 0.5 (0.04 to 5.3). B:FD (95% CI) ratios were CPAP days 1.3 (0.9 to 2.1), CPAP plus supplementary oxygen days 1.2 (0.7 to 1.9). B:FD (95% CI) ORs were death 2.3 (0.2 to 28), ventilation 2.1 (0.5 to 9), nasal excoriation 1.2 (0.2 to 8) and pneumothorax 2.4 (0.2 to 26). Conclusions In a resource-limited setting we found B CPAP equivalent to FD CPAP in the total number of days receiving CPAP within a margin of 2days.