摘要

The aim of the present study was to describe and validate a novel method able to fast, accurately and with a high precision determine the moisture content of biological materials ranging 10% to 70%. A calibrated Dual X-ray photon absorptiometry technique is validated against moisture content in wood chips. Conventional criteria such as the Pearson correlation coefficient, the root mean square errors of calibration and coefficient of variation were used for the evaluation of the method. The method is independent of the thickness and density of the object but dependent of the effective atomic number of the material. The results shows that this photon absorptiometric method gives accurate (<2%) and precise (<1%) results for moisture content determination in wood chips.

  • 出版日期2010-12