The estimation of actual crop evapotranspiration (ETc) from any given land cover or crop type is important for irrigation water management and agricultural water consumption analysis. The main parameter used for such estimations is the crop coefficient (K-c). Spectral reflectance indices, such as the normalized difference vegetation index (NDVI) and the crop coefficient of a specific crop or pasture canopy are important indicators of 'vigour', namely the photosynthetic activity and rate of biomass accumulation. Measuring both parameters simultaneously, with a view to understanding how they interact, or for creating optical, surrogate indicators of K-c is very difficult because K-c itself is difficult to measure. In this study a portable enclosed chamber was used to measure ET, of a pasture and subsequently calculated K-c from reference evapotranspiration (ETc) data derived from a nearby automatic weather station (AWS). Calibration of the chamber confirms the suitability of the device to measure the amount of water vapour produced by local plant evapotranspiration, producing a calibration factor (C) close to 1 (C = 1.02, R-2 = 0.87). The coincident NDVI values were measured using a portable active optical sensor. In a test involving a pasture (Festuca arundinacea var. Dovey) at two different stages of growth in two consecutive growing seasons, the NDVI and crop coefficients were observed to be strongly correlated (R-2 = 0.80 and 0.77, respectively). A polynomial regression (R-2 = 0.84) was found to be the best fit for the combined, multi-temporal KC-NDVI relationship. The main advantages of this method include the suitability of operating at a smaller scale (< 1 m(2)), in real time and repeatability.