摘要
An improved 129I-AMS measurement method has been developed on the 12UD tandem accelerator at the University of Tsukuba. These improvements include making a deflector and installing it behind the 120° magnet to mend the measurement of 127I current and 129I count. With MoQz molecular pilot beam to stabilize accelerator terminal voltage, the terminal voltage is kept stably within 0.1% accuracy. Combining with second stripper, this method is proved to be effective. The level of background is 7.83 10-13 for AgI reagent sample. The uncertainty of the measurement is under 7% for a series of samples which 129I/127I ratio are between 4.92 × 10-10 and 0.28 × 10-10.
- 出版日期2007