A 345 mu W Multi-Sensor Biomedical SoC With Bio-Impedance, 3-Channel ECG, Motion Artifact Reduction, and Integrated DSP

作者:Van Helleputte Nick*; Konijnenburg Mario; Pettine Julia; Jee Dong Woo; Kim Hyejung; Morgado Alonso; Van Wegberg Roland; Torfs Tom; Mohan Rachit; Breeschoten Arjan; de Groot Harmke; Van Hoof Chris; Yazicioglu Refet Firat
来源:IEEE Journal of Solid-State Circuits, 2015, 50(1): 230-244.
DOI:10.1109/JSSC.2014.2359962

摘要

This paper presents a MUlti-SEnsor biomedical IC (MUSEIC). It features a high-performance, low-power analog front-end (AFE) and fully integrated DSP. The AFE has three biopotential readouts, one bio-impedance readout, and support for general-purpose analog sensors The biopotential readout channels can handle large differential electrode offsets (+/-400 mV), achieve high input impedance (>500 M Omega), low noise (<620 nVrms in 150 Hz), and large CMRR (>110 dB) without relying on trimming while consuming only 31 mu W/channel. In addition, fully integrated real-time motion artifact reduction, based on simultaneous electrode-tissue impedance measurement, with feedback to the analog domain is supported. The bio-impedance readout with pseudo-sine current generator achieves a resolution of 9.8 m Omega/root Hz while consuming just 58 mu W/channel. The DSP has a general purpose ARM Cortex M0 processor and an HW accelerator optimized for energy-efficient execution of various biomedical signal processing algorithms achieving 10x or more energy savings in vector multiply-accumulate executions.

  • 出版日期2015-1