A simple, fast, inexpensive capillary zone electrophoresis method for the separation and determination of tiapride hydrochloride and its two related impurities in pharmaceutical formulations has been developed and validated. The successful separation of these compounds was achieved in less than 3 min using a fused silica capillary and photodiode array detector at 218 nm. The best conditions were obtained using a 10 mM sodium tetraborate (pH 8.0) as the running buffer. The linear responses covered the ranges from 1.0 to 100 mu g mL(-1) (R = 0.9989) for tiapride hydrochloride. The detection (LOD) and quantitation limits (LOQ) for tiapride hydrochloride were 2.7 and 9.0 mu g mL(-1), respectively. The intra- and inter-day relative standard deviations for migration times and peak areas were less than 0.47 and 5.7%, respectively. The method was validated for the determination of tiapride hydrochloride in commercial tablets.