Analysis of Samples Stored as Individual Plugs in a Capillary by Electrospray Ionization Mass Spectrometry

作者:Pei Jian; Li Qiang; Lee Mike S; Valaskovic Gary A; Kennedy Robert T*
来源:Analytical Chemistry, 2009, 81(15): 6558-6561.
DOI:10.1021/ac901172a

摘要

Droplets or plugs within multiphase microfluidic systems have rapidly gained interest as a way to manipulate samples and chemical reactions on the femtoliter to microliter scale. Chemical analysis of the plugs remains a challenge. We have discovered that nanoliter plugs of sample separated by air or oil can be analyzed by electrospray ionization mass spectrometry when pumped directly into a fused silica nanospray emitter tip. Using leucine-enkephalin in methanol and 1% acetic acid in water (50:50 v:v) as a model sample, we found carry-over between plugs was < 0.1% and relative standard deviation of signal for a series of plugs was 3%. Detection limits were 1 nM. Sample analysis rates of 0.8 Hz were achieved by pumping 13 nL samples separated by 3 mm long air gaps in a 75 mu m inner diameter tube. Analysis rates were limited by the scan time of the ion trap mass spectrometer. The system provides a robust, rapid, and information-rich method for chemical analysis of sample in segmented flow systems.

  • 出版日期2009-8-1