摘要

Objective: The acceptable noise level (ANL) test is used for quantification of the amount of background noise subjects accept when listening to speech. This study investigates Danish hearing-aid users%26apos; ANL performance using Danish and non-semantic speech signals, the repeatability of ANL, and the association between ANL and outcome of the international outcome inventory for hearing aids (IOI-HA). Design: ANL was measured in three conditions in both ears at two test sessions. Subjects completed the IOI-HA and the ANL questionnaire. Study sample: Sixty-three Danish hearing-aid users; fifty-seven subjects were full time users and 6 were part time/non users of hearing aids according to the ANL questionnaire. Results: ANLs were similar to results with American English speech material. The coefficient of repeatability (CR) was 6.5-8.8 dB. IOI-HA scores were not associated to ANL. Conclusions: Danish and non-semantic ANL versions yield results similar to the American English version. The magnitude of the CR indicates that ANL with Danish and non-semantic speech materials is not suitable for prediction of individual patterns of future hearing-aid use or evaluation of individual benefit from hearing-aid features. The ANL with Danish and non-semantic speech materials is not related to IOI-HA outcome.

  • 出版日期2012-9