Dosimetric consequences of pencil beam width variations in scanned beam particle therapy

作者:Chanrion M A*; Ammazzalorso F; Wittig A; Engenhart Cabillic R; Jelen U
来源:Physics in Medicine and Biology, 2013, 58(12): 3979-3993.
DOI:10.1088/0031-9155/58/12/3979

摘要

Scanned ion beam delivery enables the highest degree of target dose conformation attainable in external beam radiotherapy. Nominal pencil beam widths (spot sizes) are recorded during treatment planning system commissioning. Due to changes in the beam-line optics, the actual spot sizes may differ from these commissioning values, leading to differences between planned and delivered dose. The purpose of this study was to analyse the dosimetric consequences of spot size variations in particle therapy treatment plans. For 12 patients with skull base tumours and 12 patients with prostate carcinoma, scanned-beam carbon ion and proton treatment plans were prepared and recomputed simulating spot size changes of (1) +/- 10% to simulate the typical magnitude of fluctuations, (2) +/- 25% representing the worst-case scenario and (3) +/- 50% as a part of a risk analysis in case of fault conditions. The primary effect of the spot size variation was a dose deterioration affecting the target edge: loss of target coverage and broadening of the lateral penumbra (increased spot size) or overdosage and contraction of the lateral penumbra (reduced spot size). For changes %26lt;= 25%, the resulting planning target volume mean 95%-isodose line coverage (CI-95%) deterioration was ranging from negligible to moderate. In some cases changes in the dose to adjoining critical structures were observed.

  • 出版日期2013-6-21