A Comparison of EGFR Mutation Testing Methods in Lung Carcinoma: Direct Sequencing, Real-time PCR and Immunohistochemistry

作者:Angulo Barbara*; Conde Esther; Suarez Gauthier Ana; Plaza Carlos; Martinez Rebeca; Redondo Pilar; Izquierdo Elisa; Rubio Viqueira Belen; Paz Ares Luis; Hidalgo Manuel; Lopez Rios Fernando
来源:PLos One, 2012, 7(8): e43842.
DOI:10.1371/journal.pone.0043842

摘要

The objective of this study is to compare two EGFR testing methodologies (a commercial real-time PCR kit and a specific EGFR mutant immunohistochemistry), with direct sequencing and to investigate the limit of detection (LOD) of both PCR-based methods. We identified EGFR mutations in 21 (16%) of the 136 tumours analyzed by direct sequencing. Interestingly, the Therascreen EGFR Mutation Test kit was able to characterize as wild-type one tumour that could not be analyzed by direct sequencing of the PCR product. We then compared the LOD of the kit and that of direct sequencing using the available mutant tumours. The kit was able to detect the presence of a mutation in a 1% dilution of the total DNA in nine of the 18 tumours (50%), which tested positive with the real-time quantitative PCR method. In all cases, EGFR mutation was identified at a dilution of 5%. Where the mutant DNA represented 30% of the total DNA, sequencing was able to detect mutations in 12 out of 19 cases (63%). Additional experiments with genetically defined standards (EGFR Delta E746-A750/+ and EGFR L858R/+) yielded similar results. Immunohistochemistry (IHC) staining with exon 19-specific antibody was seen in eight out of nine cases with E746-A750del detected by direct sequencing. Neither of the two tumours with complex deletions were positive. Of the five L858R-mutated tumours detected by the PCR methods, only two were positive for the exon 21-specific antibody. The specificity was 100% for both antibodies. The LOD of the real-time PCR method was lower than that of direct sequencing. The mutation specific IHC produced excellent specificity.

  • 出版日期2012-8-27