A New Fine-Needle Aspiration System

作者:Glehr Mathias*; Leithner Andreas; Gruber Gerald; Wretschitsch Paul; Zacherl Maximilian; Kroneis Thomas; Quehenberger Franz; Windhager Reinhard
来源:Surgical Innovation, 2010, 17(2): 136-141.
DOI:10.1177/1553350610364991

摘要

Background. The main reasons for fine needle aspiration (FNA) failure are insufficient number of cells in the sample, nonrepresentative samples, and contamination of samples. The aim of this study was to measure the number of cells harvested by a new needle system with an aeration aperture (Thyrosampler) in comparison with a conventional FNA system (C-FNA). Methods. Under a double-blind setting, 30 aspirations, 15 with each system (C-FNA, Thyrosampler), were done in randomized order and recorded. Results. The median total number of cells was 59 680 cells/mL with C-FNA and 396 400 cells/mL with Thyrosampler. The needle system with the aeration aperture led to a significantly higher cell amount (564% more cells than the conventional system; P < .005) in needle aspiration biopsy. Conclusion. The new system with the vacuum release feature leads to a significantly higher cell amount in needle aspiration biopsy, which is a well-defined benefit.

  • 出版日期2010-6