摘要
The present study was conducted to assess the diagnostic accuracy of Acoustic Structure Quantification (ASQ) ultrasound software in liver biopsy of patients with liver fibrosis and cirrhosis. Eighty patients (47 +/- 14 y, 41 men) with chronic liver diseases underwent ultrasound examination of the liver and liver biopsy. In addition to the standard-care ultrasound examination, three valid gray-scale images were obtained for each patient. With the ASQ software, the average and peak values (C-m(2)) of each ultrasound gray-scale image were calculated and then compared with histologic fibrosis staging (F0-F4). No correlation was found between ASQ values and histologic fibrosis stage (p %26gt; 0.05). Areas under the curve for the diagnosis of no or mild fibrosis (F0 and F1), moderate/severe fibrosis (F2 and F3) and cirrhosis (F4) using average/peak C-m(2) values of small regions of interest were 0.46/0.43, 0.62/0.68 and 0.38/0.33. Determination of liver fibrosis with ASQ in its present form as an alternative approach to liver biopsy is too imprecise.