Assessment of aortic valve calcium load by multidetector computed tomography. Anatomical validation, impact of scanner settings and incremental diagnostic value

作者:Boulif Jamila; Gerber Bernhard; Slimani Alisson; Lazam Siham; de Meester Christophe; Pierard Sophie; Pasquet Agnes; Pouleur Anne Catherine; Vancraeynest David; El Khoury Gebrine; de Kerchove Laurent; Noirhomme Philippe; Vanoverschelde Jean Louis*
来源:Journal of Cardiovascular Computed Tomography, 2017, 11(5): 360-366.
DOI:10.1016/j.jcct.2017.07.004

摘要

Objectives: To validate aortic valve calcium (AVC) load measurements by multidetector row computed tomography (MDCT), to evaluate the impact of tube potential and slice thickness on AVC scores, to examine the accuracy of AVC load in distinguishing severe from nonsevere aortic stenosis (AS) and to investigate its effectiveness as an alternative diagnosis method when echocardiography remains inconclusive. Methods: We prospectively studied 266 consecutive patients with moderate to severe AS who underwent MDCT to measure AVC load and a comprehensive echocardiographic examination to assess AS severity. AVC load was validated against valve weight in 57 patients undergoing aortic valve replacement. The dependence of AVC scores on tube potential and slice thickness was also tested, as well as the relationship between AVC load and echocardiographic criteria of AS severity. Results: MDCT Agatston score correlated well with valve weight (r = 0.82, p < 0.001) and hemodynamic indices of AS severity (all p < 0.001). Ex-vivo Agatston scores decreased significantly with increasing tube potential and slice thickness (repeated measures ANOVA p < 0.001). Multivariate analysis identified mean gradient, the indexed effective orifice area, male gender and left ventricular outflow tract cross-sectional area as independent correlates of the in-vivo AVC load. Conclusions: MDCT-derived AVC load correlated well with valve weight and hemodynamic indices of AS severity. It also depends on tube potential and slice thickness, thus suggesting that these parameters should be standardized to optimize reproducibility and accuracy.

  • 出版日期2017-10