A novel fast screening method was developed for the determination of polychlorinated biphenyls that are constituents of the commercial mixture, Aroclor 1260, in soil matrices by gas chromatography with mass spectrometry combined with solid-phase microextraction. Nonequilibrium headspace solid-phase microextraction with a 100 m polydimethylsiloxane fiber was used to extract polychlorinated biphenyls from 0.5 g of soil matrix. The use of 2 mL of saturated potassium dichromate in 6 M sulfuric acid solution improved the reproducibility of the extractions and the mass transfer of the polychlorinated biphenyls from the soil matrix to the microextraction fiber via the headspace. The extraction time was 30 min at 100 degrees C. The percent recoveries, which were evaluated using an Aroclor 1260 standard and liquid injection, were within the range of 54.9-65.7%. Two-way extracted ion chromatogram data were used to construct calibration curves. The relative error was %26lt;+/- 15% and the relative standard deviation was %26lt;15%, which are respective measures of the accuracy and precision. The method was validated with certified soil samples and the predicted concentrations for Aroclor 1260 agreed with the certified values. The method was demonstrated to be linear from 10 to 1000 ng/g for Aroclor 1260 in dry soil.