Determination of Reynolds Shear Stress Level for Hemolysis

作者:Jhun Choon Sik; Stauffer Megan A; Reibson John D; Yeager Eric E; Newswanger Raymond K; Taylor Joshua O; Manning Keefe B; Weiss William J; Rosenberg Gerson*
来源:ASAIO Journal, 2018, 64(1): 63-69.
DOI:10.1097/MAT.0000000000000615

摘要

Reynolds shear stress (RSS) has served as a metric for the effect of turbulence on hemolysis. Forstrom (1969) and Sallam and Hwang (1984) determined the RSS threshold for hemolysis to be 50,000 and 4,000 dyne/cm(2), respectively, using a turbulent jet. Despite the order of magnitude discrepancy, the threshold by Sallam and Hwang has been frequently cited for hemolytic potential in blood pumps. We recreated a Sallam apparatus (SA) to resolve this discrepancy and provide additional data to be used in developing a more accurate hemolysis model. Hemolysis was measured over a large range of Reynolds numbers (Re) (Re = 1,000-80,000). Washed bovine red blood cells (RBCs) were injected into the free jet of phosphate buffered saline, and hemolysis was quantified using a percent hemolysis, H-p = h (100 -hematocrit [HCT])/Hb, where h (mg/dl) is free hemoglobin and Hb (mg/dl) is total hemoglobin. Reynolds shear stress was calculated using two-dimensional laser Doppler velocimetry. Reynolds shear stress of >= 30,000 dyne/cm(2) corresponding to Re of >= 60,000 appeared to cause hemolysis (p < 0.05). This RSS is an order of magnitude greater than the RSS threshold that Sallam and Hwang suggested, and it is similar to Forstrom's RSS threshold. This study resolved a long-standing uncertainty regarding the critical values of RSS for hemolysis and may provide a foundation for a more accurate hemolysis model.

  • 出版日期2018-2