摘要
Objective To develop a method for the assessment of small field of view (SFOV) gamma cameras using a novel phantom designed to simulate the localization of sentinel nodes in the presence of a high-activity injection site. Materials and methods The phantom consisted of a cube-shaped acrylic glass support frame. Sixteen acrylic glass plates and nine bars were stacked within the frame to allow a variable configuration of the simulated node depth and node-to-injection site separation. Syringes filled with Tc-99m were used to simulate activity at the injection site and node. Scintigraphic imaging was carried out and the images were assessed subjectively and quantitatively through calculation of the contrast-to-noise ratio. The detection performance of an SFOV gamma camera was then compared with that of a large field of view gamma camera. Results The detectability studies showed that the SFOV gamma camera could detect low activity in nodes by visual examination of images and with contrast-to-noise ratios ranging from 3 to 62. In particular, the phantom showed the limits of node detection using an SFOV gamma camera over activity ratios less than 1 : 100 and at depths below 45mm with 25mm of node-to-injection site separation. Visual subjective assessment of images acquired under the same conditions showed that the SFOV gamma camera was superior to a large field of view camera for the detection of nodes at a node-to-injection site separation of 25 mm. Conclusion A low-cost phantom has been designed and fabricated that provides a versatile method for the assessment of SFOV gamma cameras intended for sentinel node imaging.
- 出版日期2015-11