A new mathematical model of wrist pulse waveforms characterizes patients with cardiovascular disease - A pilot study

作者:He, Dianning; Wang, Lu; Fan, Xiaobing; Yao, Yang; Geng, Ning; Sun, Yingxian; Xu, Lisheng*; Qian, Wei
来源:Medical Engineering & Physics, 2017, 48: 142-149.
DOI:10.1016/j.medengphy.2017.06.022

摘要

The purpose of this study was to analyze and compare a series of measured radial pulse waves as a function of contact pressure for young and old healthy volunteers, and old patients with cardiovascular disease. The radial pulse waves were detected with a pressure sensor and the contact pressure of the sensor was incremented by 20 gf during the signal acquisition. A mathematical model of radial pulse waveform was developed by using two Gaussian functions modulated by radical functions and used to fit the pulse waveforms. Then, a ratio of area (r(A)) and a ratio of peak height (r(PH)) between percussion wave and dicrotic wave as a function of contact pressure were calculated based on fitted parameters. The results demonstrated that there was a maximum for waveform peak height, a minimum for r(A) (r(A)(min)) and a minimum for r(PH) (r(PH)(min)) appeared as contact pressure varied. On average, older patients had higher peak amplitude and a significantly smaller r(A)(min) (p < 0.001) and r(PH)(min) (p < 0.02) than the young and old volunteers. The r(A)(min) and r(PH)(min) calculated with the mathematical model had moderate to strong positive linear correlations (r = 0.66 to 0.84, p < 0.006) with those directly calculated without the model. The receiver operating characteristic (ROC) analysis showed that the r(A)(min) calculated with the model and the contact pressure measured at the r(A)(min) had good diagnostic accuracy to distinguish healthy volunteers vs. diseased patients. Therefore, using the mathematical model to quantitatively analyze the radial pulse waveforms as a function of contact pressure could be useful in the diagnosis of cardiovascular diseases.