Accuracy of T1 measurement with 3-D Look-Locker technique for dGEMRIC

作者:Li, Wei*; Scheidegger, Rachel; Wu, Ying; Vu, Anthony; Prasad, Pottumarthi V.
来源:Journal of Magnetic Resonance Imaging, 2008, 27(3): 678-682.
DOI:10.1002/jmri.21244

摘要

Purpose: To validate the accuracy of T1 measurement by three-dimensional Look-Locker method (3D LL) for delayed gadolinium-enhanced MRI of cartilage (dGEMRIC) of human subjects with and without osteoarthritis (OA), as compared with two-dimensional inversion recovery fast spin-echo (213 IR-FSE) technique. @@@ Materials and Methods: MR sagittal images of the knees were acquired for T I mapping in 29 subjects with standard 2D IR-FSE and 3D LL sequences 90-135 min following administration of 0.2 mmol/kg Gd-DTPA(2-). T1 maps of femoral and tibial cartilage were generated using custom software. Comparisons in T1 values between the two techniques were performed using regression analysis. @@@ Results: Good agreement in T1 values between 2D IR-FSE and 3D LL was observed (R values of 0.90, and 0.85, and 0.86 for all, OA, and control subjects, respectively) when acquired within 15 min. @@@ Conclusion: The 3D LL sequence provides accurate T1 estimates of articular cartilage with advantages of entire joint coverage, shorter acquisition time, and a wide range of inversion times sampled within a single acquisition.