P>Current laboratory standards from Clinical Laboratory Standards Institute (CLSI) and manufacturer's (Becton Dickinson) data indicate that under-filling K(2)EDTA blood collection tubes can result in erroneous hematology values. To accommodate under-filled tubes and reduce collection volumes while optimizing our automation, we explored the acceptable limit of under-filled tubes for hematology values. We collected 8.0 ml of blood from 30 normal adult volunteers. Each donation was aliquoted in the following volumes: 4.0, 2.0, 1.0, 0.5 ml x 2. These samples were analyzed within 1 h of blood collection on Sysmex XE-2100 (Sysmex America Inc., Mundelein, IL, USA) for complete blood count, reticulocyte, and white blood cell differentials. Results of the under-filled tubes were compared to those of the standard volume. The Deming regression analysis show excellent correlation for all parameters between each under-filled blood collection volume compared to a standard 4 ml volume. The Bland and Altman analysis shows good agreement between both 1.0 and 2.0 ml compared to a 4.0 ml volume. The 0.5 ml compared to a 4.0 ml volume, however, shows increased variation on many parameters. In addition all three collection volumes show negative bias compared to the standard volume for platelet count, but the difference is considered insignificant with a percent difference of 5.5%, 3.2%, and 1.5% for 0.5, 1.0, and 2.0 ml collection volume respectively. Finally for 0.5 ml collection volume we noticed a low level of false positive flagging rate for white blood cell. Acceptable complete blood count values of under-filled powdered K(2)EDTA tubes can be obtained with as little as 1.0 ml of blood.