摘要

We developed and optimized a custom dynamic Electronic Speckle Pattern Interferometer (d-ESPI) for measuring time-dependent three-dimensional (3D) surface displacements during inflation testing of ocular tissues. The 3D displacement field was resolved under dynamic loading conditions for a simplified test case of a rubber phantom and a real-world test case of the posterior sclera of a human eye. We present the optical layout and calibration procedure of the d-ESPI, and demonstrate how the displacement field can be accurately resolved for specimen deformation rates up to 6 mu m/s.

  • 出版日期2018-8