摘要

An analysis method was developed to detect trace elements in Chinese Honeylocust Fruit by using inductively coupled plasma tandem mass spectrometry (ICP-MS/MS). All samples were microwave digested by using (HNO3 + H2O2 HCl). In the MS/MS mode, the polyatomic interferences were eliminated by the utilization of O-2 mass shift, O-2 on-mass, and NH3 mass shift with O-2 or NH3 in the collision/reaction cell of the ICP-MS/MS. The analytical results of proposed method were validated by analysis of the national standard reference samples of green tea (GBW10052) and celery (GBW10048), which showed good agreement with the certified values, RSD <= 7. 8%. Furthermore, 12 trace elements (V, Cr, Co, Cu, Zn, Fe, Mn, Ni, As, Cd, Hg, and Pb) in these samples collected from Yunnan and Chongqing Wuling Mountain, were detected by proposed analyzed method. The detection limits of the analytes were in the range of 0. 29 similar to 21. 36 ng.L-1 The analytical results showed that Fe and Zn elements were rich in these wild samples, while heavy elements in these samples were lower than limitation standard of Chinese pharmacopoeia. this study can be used to estimate the trace element and quality of wild honey locust. Also, this study can provide scientific methods and data for the safety assessment and regulation of trace elements in Chinese Honeylocust Fruit.