Across-subject calibration of an instrumented glove to measure hand movement for clinical purposes

作者:Gracia Ibanez Veronica*; Vergara Margarita; Buffi James H; Murray Wendy M; Sancho Bru Joaquin L
来源:Computer Methods in Biomechanics and Biomedical Engineering, 2017, 20(6): 587-597.
DOI:10.1080/10255842.2016.1265950

摘要

Motion capture of all degrees of freedom of the hand collected during performance of daily living activities remains challenging. Instrumented gloves are an attractive option because of their higher ease of use. However, subject-specific calibration of gloves is lengthy and has limitations for individuals with disabilities. Here, a calibration procedure is presented, consisting in the recording of just a simple hand position so as to allow capture of the kinematics of 16 hand joints during daily life activities even in case of severe injured hands. across-subject gains' were obtained by averaging the gains obtained from a detailed subject-specific calibration involving 44 registrations that was repeated three times on multiple days to 6 subjects. In additional 4 subjects, joint angles that resulted from applying the across-subject calibration' or the subject-specific calibration were compared. Global errors associated with the across-subject calibration' relative to the detailed, subject-specific protocol were small (bias: 0.49 degrees; precision: 4.45 degrees) and comparable to those that resulted from repeating the detailed protocol with the same subject on multiple days (0.36 degrees; 3.50 degrees). Furthermore, in one subject, performance of the across-subject calibration' was directly compared to another fast calibration method, expressed relative to a videogrammetric protocol as a gold-standard, yielding better results.