摘要

To obtain the correct tomographic reconstruction of micron-sized materials, the nonlinear intensity attenuation of bright-field transmission electron microscopy (BF-TEM) images was analyzed as a function of the sample thickness using a high-voltage electron microscope. The intensity attenuation was precisely measured relative to the projection thickness of carbon microcoils (CMCs) at acceleration voltages of 400-1000 kV using objective apertures (OAs) with radii of 2.1-28 nm(-1). The results show that the nonlinearity is strongly dependent on the OA size and the acceleration voltage. The influence of nonlinearity on tomographic reconstructions was also examined using a specially developed 360 degrees-tilt sample holder. Sliced images of the reconstructed volumes indicated that an increase in the nonlinearity caused artificial fluctuations in the internal density of materials and inaccurate shapes of the objects in more significant cases. Conditions sufficient for reconstruction with the correct density have been estimated to be 0.67 of the minimum electron transmittance, and for reconstructions with correct shapes, 0.4. This information enables foreseeing the quality of the reconstruction from a single BF-TEM image prior to the tilt-series acquisition. As a result to demonstrate the appropriateness of these conditions, a CMC with a diameter of 3.7 mu m was reconstructed successfully; i.e. not only the shape but also the internal density were correctly reproduced using electron tomography.