Development of a sequential data correction method for isotope ratio analysis by resonance ionization mass spectrometry

作者:Noto Takuma*; Tomita Hideki; Furuta Yujin; Takamatsu Takahide; Kawarabayashi Jun; Iguchi Tetsuo; Wendt Klaus
来源:Journal of Nuclear Science and Technology, 2016, 53(2): 289-294.
DOI:10.1080/00223131.2015.1038663

摘要

A sequential data correction method using several experimental parameters of pulsed resonance ionization mass spectrometry was developed to reduce mass discrimination effects for isotope ratio analysis as well as fluctuations in the ion count rate. This correction method was applied to isotope ratio determination of stable titanium isotopes using a two-step, single color ionization scheme. A significant improvement of the precision of the measured isotopic ratio of stable titanium isotopes was demonstrated after correction with a parameter set consisting of laser power, atmospheric pressure, and laser frequency.

  • 出版日期2016-2-1