A piezoelectric-based immunosensor for high density lipoprotein particle measurement

作者:Chunta Suticha; Suk Anake Jamikorn; Chansiri Kosum; Promptmas Chamras*
来源:Analyst, 2014, 139(18): 4586-4592.
DOI:10.1039/c4an00601a

摘要

A piezoelectric-based immunosensor was developed for high density lipoprotein particle (HDL-P) measurement. Monoclonal anti-human apolipoprotein A1 antibody was used as a specific binding molecule for the major apolipoprotein of HDL-P. This sensing element was fabricated by immobilizing the anti-human apolipoprotein A1 on a 12 MHz AT-cut quartz crystal via a 3-mercaptopropionic acid (MPA) self-assembled monolayer. The frequency shift from the mass change of the antigen-antibody binding refers to the amount of HDL-P. The optimal antibody immobilization was performed to achieve the maximum potential of the antibody. The appropriate quantity and immobilization time of the antibody were 0.1 mg ml(-1) and 90 minutes, respectively. The immobilized antibody in the HDL-P immunosensor accomplished perfect binding with HDL-P within 60 minutes. The dose-response curve for HDL-P showed a linear response from 0.21 to 2.50 mg protein per ml equivalent to 0.40 x 10(10) to 3.65 x 10(10) particles per mu l without significant interference from other lipoproteins. The intra- and interassay imprecision (CV) were 7.8 and 18.5%, respectively. The analytical accuracy of this measurement was 96.29-96.31%. The HDL-P concentration obtained from the sensor revealed a 2.05 mg protein per ml with 0.26 mg protein per ml of expanded uncertainty at the 95% confidence level. This immunosensor gave an assay result which correlated with the homogeneous enzymatic colorimetric assay (R-2 = 0.902).

  • 出版日期2014-9-21