Purpose: It is unclear that spatial accuracy can reflect the impact of deformed dose distribution. In this study, we used dosimetric parameters to compare an in-house deformable image registration (DIR) system using NiftyReg, with two commercially available systems, MIM Maestro (MIM) and Velocity AI (Velocity). Methods: For 19 non-small-cell lung cancer patients, the peak inspiration (0%)-4DCT images were deformed to the peak expiration (50%)-4DCT images using each of the three DIR systems, which included computation of the deformation vector fields (DVF). The 0%-gross tumor volume (GTV) and the 0%-dose distribution were also then deformed using the DVFs. The agreement in the dose distributions for the GTVs was evaluated using generalized equivalent uniform dose (gEUD), mean dose (D-mean), and three-dimensional (3D) gamma index (criteria: 3 mm/3%). Additionally, a Dice similarity coefficient (DSC) was used to measure the similarity of the GTV volumes. Results: D-mean and gEUD demonstrated good agreement between the original and deformed dose distributions (differences were generally less than 3%) in 17 of the patients. In two other patients, the Velocity system resulted in differences in gEUD of 50.1% and 29.7% and in D-mean of 11.8% and 4.78%. The gamma index comparison showed statistically significant differences for the in-house DIR vs. MIM, and MIM vs. Velocity. Conclusions: The finely tuned in-house DIR system could achieve similar spatial and dose accuracy to the commercial systems. Care must be taken, as we found errors of more than 5% for D-mean and 30% for gEUD, even with a commercially available DIR tool.