Microwave ablation of ex vivo human liver and colorectal liver metastases with a novel 14.5 GHz generator

作者:Jones Robert P*; Kitteringham Neil R; Terlizzo Monica; Hancock Christopher; Dunne Declan; Fenwick Stephen W; Poston Graeme J; Ghaneh Paula; Malik Hassan Z
来源:International Journal of Hyperthermia, 2012, 28(1): 43-54.
DOI:10.3109/02656736.2011.610428

摘要

Purpose: This study assessed the relationship between time, power and ablation size using a novel high-frequency 14.5GHz microwave applicator in ex vivo human hepatic parenchyma and colorectal liver metastases. Previous examination has demonstrated structurally normal but non-viable cells within the ablation zone. This study aimed to further investigate how ablation affects these cells, and to confirm non-viability. %26lt;br%26gt;Materials and methods: Ablations were performed in ex vivo human hepatic parenchyma and tumour for a variety of time (10-180 s) and power (10-50W) settings. Histological examination was performed to assess cellular anatomy, whilst enzyme histochemistry was used to confirm cellular non-viability. Transmission electron microscopy was used to investigate the subcellular structural effects of ablation within these fixed cells. Preliminary proteomic analysis was also performed to explore the mechanism of microwave cell death. %26lt;br%26gt;Results: Increasing time and power settings led to a predictable and reproducible increase in size of ablation. At 50W and 180s application, a maximum ablation diameter of 38.8mm (+/- 1.3) was produced. Ablations were produced rapidly, and at all time and power settings ablations remained spherical (longest: shortest diameter %26lt; 1.2). Routine histological analysis using haematoxylin-eosin (H%26 E) confirmed well preserved cellular anatomy despite ablation. Transmission electron microscopy demonstrated marked subcellular damage. Enzyme histochemistry showed complete absence of viability in ablated tissue. %26lt;br%26gt;Conclusions: Large spherical ablation zones can be rapidly and reproducibly achieved in ex vivo human hepatic parenchyma and colorectal liver metastases using a 14.5GHz microwave generator. Despite well preserved cellular appearance, ablated tissue is non-viable.

  • 出版日期2012