A Robotic Ultrasound Scanner for Automatic Vessel Tracking and Three-Dimensional Reconstruction of B-Mode Images

作者:Merouche Samir*; Allard Louise; Montagnon Emmanuel; Soulez Gilles; Bigras Pascal; Cloutier Guy
来源:IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 2016, 63(1): 35-46.
DOI:10.1109/TUFFC.2015.2499084

摘要

Locating and evaluating the length and severity of a stenosis is very important for planning adequate treatment of peripheral arterial disease (PAD). Conventional ultrasound (US) examination cannot provide maps of entire lower limb arteries in 3-D. We propose a prototype 3D-US robotic system with B-mode images, which is nonionizing, noninvasive, and is able to track and reconstruct a continuous segment of the lower limb arterial tree between the groin and the knee. From an initialized cross-sectional view of the vessel, automatic tracking was conducted followed by 3D-US reconstructions evaluated using Hausdorff distance, cross-sectional area, and stenosis severity in comparison with 3-D reconstructions with computed tomography angiography (CTA). A mean Hausdorff distance of 0.97 +/- 0.46 mm was found in vitro for 3D-US compared with 3D-CTA vessel representations. To evaluate the stenosis severity in vitro, 3D-US reconstructions gave errors of 3%-6% when compared with designed dimensions of the phantom, which are comparable to 3D-CTA reconstructions, with 4%-13% errors. The in vivo system's feasibility to reconstruct a normal femoral artery segment of a volunteer was also investigated. These results encourage further ergonomic developments to increase the robot's capacity to represent lower limb vessels in the clinical context.

  • 出版日期2016-1