Determination of the variation of the fluorescence line positions of ruby, strontium tetraborate, alexandrite, and samarium-doped yttrium aluminum garnet with pressure and temperature

作者:Raju Selva Vennila*; Zaug Joseph M; Chen Bin; Yan Jinyuan; Knight Jason W; Jeanloz Raymond; Clark Simon M
来源:Journal of Applied Physics, 2011, 110(2): 023521.
DOI:10.1063/1.3608167

摘要

The pressure and temperature dependent fluorescence line-shift of strontium tetraborate has been measured concurrently with x-ray diffraction from the pressure standards sodium chloride or gold. Temperature was found to have a small effect on the fluorescence line-shift under pressure. We found a maximum pressure uncertainty of +/- 1.8 GPa at 25 GPa (7.2%) and 857 K when making no temperature correction. The fluorescence line-shifts for ruby, Alexandrite, and samarium-doped yttrium aluminum garnet were also determined, using our strontium tetraborate calibration to determine pressure and a thermocouple to measure temperature. Fluorescence measurements were extended up to 800 K for ruby and Alexandrite. Temperature was found to have a small effect on the fluorescence line-shift of samarium-doped yittrium aluminum garnet. We found a maximum uncertainty of +/- 2.7 GPa at 25 GPa (11.1%) and 857 K when no temperature correction was applied. We determined equations relating to the fluorescence line position from these data, which include a cross derivative term to account for the combined effect of pressure and temperature. We present a method to independently determine pressure and/or temperature from combined fluorescence line-shift measurements of a pair of optical sensors.

  • 出版日期2011-7-15