摘要

BACKGROUND: Extravascular catheter placement during implantation of totally implantable venous access port (TIAP) is a rare complication. METHODS: Case report. RESULTS: We report a case of a 48-year-old woman with a normal postoperative chest x-ray developing a massive fluidopneumothorax after first application of chemotherapy via TIAP. Intravenous catheter placement was not confirmed by aspiration of blood after TIAP implantation. Postoperative chest X-ray was performed describing the catheter to be in correct position. CONCLUSIONS: Aspiration of blood should be done directly after TIAP implantation to prove intravascular position of catheter. Postoperative chest X-ray without a contrast medium fails to detect extravascular catheter placement.

  • 出版日期2012-8

全文