The dose-to-curie (DTC) conversion method has been known that there could be extremely high uncertainty associated with establishing the radioactivity of gamma emitters in a drum. However, the DTC conversion method is still an effective assay method to calculate the radioisotope inventory because of the simple and easy procedures to be applied. In order to make the DTC conversion method practical, numerous assumptions and limitations placed on its use. These assumptions and limitations are related to the dose rate measurement and the relative abundance of gamma emitters in a drum. However, these two variables were generally obtained from the different detection mechanisms even using the different radwaste each other. Unfortunately, that expanded the limitation of using the DTC conversion method. In order to obtain two variables in a drum to be assayed at once, the dose conversion factor for a NaI(Tl) detector was first calculated from the MCNP code. The pulse height spectrum from a simulated drum inserted into a standard source was measured by a NaI(Tl) detector, and then, two variables were calculated from the dose conversion factor and the net count rate of detected gamma emitters in the pulse height spectrum.