Cardiac monitoring of ventricular assist devices (VADs) is important for detecting heart failure risks, such as critical arrhythmia and ventricular fibrillation, and for supplying data that are useful for hemodynamic control. Specifically, impedance cardiograms (ICGs) are especially beneficial because they have no effect on the tissue or organs and can monitor various parameters simultaneously, including the heart rate and heart contractions. In this article, we measured impedance changes in porcine left ventricles using electrodes placed around the inlet and outlet cannulae of the VAD. The measured left ventricular impedance (LVI) waveform changes are caused by heart movements, such as cardiac muscle contraction and changes in blood volume as a result of heart filling and emptying. In contrast to other impedance measurements, LVI is less affected by the movement of other organs. Using a porcine model, LVIs were measured and compared with blood flow data measured with an ultrasound blood flowmeter. The ICG showed the same frequency as the animal%26apos;s heart rate, and their amplitudes were closely related to cardiac output (CO). However, the waveform differed from other vital signs, such as CO, electrocardiogram, and blood pressure. Ultrasound images were used to explain the impedance waveform. In the ultrasound images, we obtained the shape and size of the animal%26apos;s heart and calculated the predicted impedance data. We then compared these to the actual measured data. These results show that the impedance signal contains detailed information on heart rate and CO; these results were unaffected by the cannulae or VAD perfusion.