Automated Off-Line Respiratory Event Detection for the Study of Postoperative Apnea in Infants

作者:Aoude Ahmed A; Kearney Robert E; Brown Karen A; Galiana Henrietta L; Robles Rubio Carlos A*
来源:IEEE Transactions on Biomedical Engineering, 2011, 58(6): 1724-1733.
DOI:10.1109/TBME.2011.2112657

摘要

Previously, we presented automated methods for thoraco-abdominal asynchrony estimation and movement artifact detection in respiratory inductance plethysmography (RIP) signals. This paper combines and improves these methods to give a method for the automated, off-line detection of pause, movement artifact, and asynchrony. Simulation studies demonstrated that the new combined method is accurate and robust in the presence of noise. The new procedure was successfully applied to cardiorespiratory signals acquired postoperatively from infants in the recovery room. A comparison of the events detected with the automated method to those visually scored by an expert clinician demonstrated a higher agreement (kappa = 0.52) than that amongst several human scorers (kappa = 0.31) in a clinical study [1]. The method provides the following advantages: first, it is fully automated; second, it is more efficient than visual scoring; third, the analysis is repeatable and standardized; fourth, it provides greater agreement with an expert scorer compared to the agreement between trained scorers; fifth, it is amenable to online detection; and lastly, it is applicable to uncalibrated RIP signals. Examples of applications include respiratory monitoring of postsurgical patients and sleep studies.

  • 出版日期2011-6