A newly designed big cup nitinol stent for gastric outlet obstruction

作者:Shi, Ding; Liao, Sheng-Hui; Geng, Jian-Ping*
来源:World Journal of Gastroenterology, 2010, 16(33): 4206-4209.
DOI:10.3748/wjg.v16.i33.4206

摘要

AIM: To find out whether a newly designed big cup nitinol stent is suitable for treatment of patients with gastric outlet obstruction resulting from gastric cancer. METHODS: The new stent is composed of a proximal big cup segment (20 mm in length and 48-55 mm in diameter), a middle part (60 mm in length and 20 mm in diameter) covered by a polyethylene membrane and a distal sphericity (20 mm in length and 28 mm in diameter). Half of the proximal big cup segment is also covered by a polyethlene membrane, which is adjacent to the middle part of the stent. The stent is preloaded in a 6.0-mm-diameter introducer system. Thirteen patients with gastric outlet obstruction resulting from gastric cancer received the new stents under endoscopic and fluoroscopic guidance. RESULTS: Technical success was achieved in 12 of 13 (92.3%) patients. Among the 12 patients in whom endoscopic stent was placed successfully, the clinical success rate was 91.7% during a follow-up of average 6.5 mo. During the first month follow-up, the migration rate was 0%, recurrent obstruction 0% and gastric bleeding 8.3%. During the follow-up between 2-12 mo, no migration, recurrent obstruction and gastric bleeding occurred. CONCLUSION: The proximal big cup segment seems to be effective and promising for technical efficacy, clinical outcome, and preventing migration and tumor ingrowth and increasing the emptying rate of sinus ventriculi.