Development of a computer-based automated pure tone hearing screening device: a preliminary clinical trial

作者:Gan Kok Beng*; Azeez Dhifaf; Umat Cila; Ali Mohd Alauddin Mohd; Wahab Noor Alaudin Abdul; Mukari Siti Zamratol Mai Sarah
来源:Biomedizinische Technik / Biomedical Engineering, 2012, 57(5): 323-332.
DOI:10.1515/bmt-2011-0086

摘要

Hearing screening is important for the early detection of hearing loss. The requirements of specialized equipment, skilled personnel, and quiet environments for valid screening results limit its application in schools and health clinics. This study aimed to develop an automated hearing screening kit (auto-kit) with the capability of real-time noise level monitoring to ensure that the screening is performed in an environment that conforms to the standard. The auto-kit consists of a laptop, a 24-bit resolution sound card, headphones, a microphone, and a graphical user interface, which is calibrated according to the American National Standards Institute S3.6-2004 standard. The auto-kit can present four test tones (500, 1000, 2000, and 4000 Hz) at 25 or 40 dB HL screening cut-off level. The clinical results at 40 dB HL screening cut-off level showed that the auto-kit has a sensitivity of 92.5% and a specificity of 75.0%. Because the 500 Hz test tone is not included in the standard hearing screening procedure, it can be excluded from the auto-kit test procedure. The exclusion of 500 Hz test tone improved the specificity of the auto-kit from 75.0% to 92.3%, which suggests that the auto-kit could be a valid hearing screening device. In conclusion, the auto-kit may be a valuable hearing screening tool, especially in countries where resources are limited.

  • 出版日期2012

全文