Application of Gafchromic film in the study of dosimetry methods in CT phantoms

作者:Martin C J*; Gentle D J; Sookpeng S; Loveland J
来源:Journal of Radiological Protection, 2011, 31(4): 389-409.
DOI:10.1088/0952-4746/31/4/001

摘要

Gafchromic film has been used for measurement of computed tomography (CT) dose distributions within phantoms. The film was calibrated in the beam from a superficial therapy unit and the accuracy confirmed by comparison with measurements with a 20 mm long ionisation chamber. The results have been used to investigate approaches to CT dosimetry. Dose profiles were recorded within standard CT head and body phantoms and scatter tail data fitted to exponential functions and extrapolated to predict dose levels in longer phantoms. The data have been used to simulate both CT dose index (CTDI) measurements with ionisation chambers of differing length and measurements of cumulative doses with a 20 mm chamber for scans of varying length. The results show that the length of a pencil ionisation chamber is the most significant factor affecting measurements of weighted CTDI (CTDI(w)) and a 100 mm chamber would record 50-61% of the dose measured with a 450 mm one. The cumulative dose measured at the centre of a 150 mm long body phantom records over 70% of the equilibrium dose from a helical scan of a longer phantom. For routine CT dosimetry tests, the determination of correction factors could allow measurements with a 100 mm chamber to be used to derive the CTDI that would be recorded with a longer chamber, and cumulative doses measured with a 20 mm chamber in shorter phantoms to be used to calculate equilibrium doses for helical scans.

  • 出版日期2011-12