摘要

A chemiluminescence (CL) detection method for the determination of sulfite using the reaction of Ru(bipy)(3)(2+) (bipy = 2,2'-bipyridyl)-SO32--KBrO3 is described. The concentration of sulfite is proportional to the CL intensity in the range 2.5 x 10(-8)-9.5 x 10(-5) mol l(-1). The limit of detection is 3.8 x 10(-9) mol l(-1) and the relative standard deviation is 4.6% for 5 x 10(-5) mol l(-1) sulfite solution with nine repeated measurements. This method was successfully applied to the determination of sulfite in sugar and sulfur dioxide in air by using triethanolamine as the absorbent material.