摘要

Optical coherence tomography (OCT) with ultrahigh axial resolution was achieved by the super-continuum generated by coupling femtosecond pulses from a commercial Ti: sapphire laser into an air-silica microstructure fiber. The visible spectrum of the super-continuum from 450 to 700 nm centered at 540 nm can be generated. A free-space axial OCT resolution of 0.64 mu m was achieved. The sensitivity of OCT system was 108 dB with incident light power 3 mW at sample, only 7dB below the theoretical limit. Subcellular OCT imaging was also demonstrated, showing great potential for biomedical application.