In this paper, we discuss approaches to prepare solid samples for X-ray fluorescence spectrometry (XRF). Although XRF can be used to analyze major and minor elements in various solid samples including powders and grains without dissolution techniques, to obtain reliable XRF results, the prepared sample must meet certain criteria related to homogeneity, particle size, flatness, and thickness. The conditions are defined by the analytical depth of fluorescent X-rays from analytes, and the analytical depth can be estimated from the X-ray absorption related to the energy of each X-ray and the composition and density of the sample. For example, when the sample flatness and particle size are less than the analytical depth and the sample possesses homogeneity within a depth less than the analytical depth, the XRF results are representative of the entire sample. Furthermore, an appropriate sample thickness that is larger than the analytical depth or constant can prevent changes in fluorescent X-ray intensity with variations in sample thickness. To obtain accurate and reproducible measurements, inhomogeneous solid samples must be pulverized, homogenized, and prepared as loose powder, powder pellets, or glass beads. This paper explains the approaches used to prepare solid samples for XRF analysis based on the analytical depths of fluorescent X-rays.