Acoustic characterization of echogenic liposomes: Frequency-dependent attenuation and backscatter

作者:Kopechek Jonathan A*; Haworth Kevin J; Raymond Jason L; Mast T Douglas; Perrin Stephen R Jr; Klegerman Melvin E; Huang Shaoling; Porter Tyrone M; McPherson David D; Holland Christy K
来源:Journal of the Acoustical Society of America, 2011, 130(5): 3472-3481.
DOI:10.1121/1.3626124

摘要

Ultrasound contrast agents (UCAs) are used clinically to aid detection and diagnosis of abnormal blood flow or perfusion. Characterization of UCAs can aid in the optimization of ultrasound parameters for enhanced image contrast. In this study echogenic liposomes (ELIPs) were characterized acoustically by measuring the frequency-dependent attenuation and backscatter coefficients at frequencies between 3 and 30 MHz using a broadband pulse-echo technique. The experimental methods were initially validated by comparing the attenuation and backscatter coefficients measured from 50-mu m and 100-mu m polystyrene microspheres with theoretical values. The size distribution of the ELIPs was measured and found to be polydisperse, ranging in size from 40 nm to 6 mu m in diameter, with the highest number observed at 65 nm. The ELIP attenuation coefficients ranged from 3.7 +/- 1.0 to 8.0 +/- 3.3 dB/cm between 3 and 25 MHz. The backscatter coefficients were 0.011 +/- 0.006 (cm str)(-1) between 6 and 9 MHz and 0.023 +/- 0.006 (cm str)(-1) between 13 and 30 MHz. The measured scattering-to-attenuation ratio ranged from 8% to 22% between 6 and 25 MHz. Thus ELIPs can provide enhanced contrast over a broad range of frequencies and the scattering properties are suitable for various ultrasound imaging applications including diagnostic and intravascular ultrasound.

  • 出版日期2011-11