Advanced interlocking systems to improve heavy-load-bearing characteristics of flexible intramedullary nailing

作者:Berger Leopold*; Eichler Johannes; Ryll E Jonathan S; Fischerauer Stefan; Raschke Michael J; Kolbitsch Andreas; Castellani Christoph; Weinberg Annelie Martina
来源:Materials Science and Engineering: C , 2016, 68: 358-365.
DOI:10.1016/j.msec.2016.05.122

摘要

Flexible intramedullary nailing (FIN) is a minimally invasive and widespread standard method for osteosynthesis of pediatric long bone fractures. In the case of unstable fractures of the lower extremity, interlocking systems need to be used to prevent axial shortening and subsequent perforation of the nail at its insertion site. In the present study, four different screw-fixed interlocking systems for FINs (Hofer TwinPlug with two 3-mm titanium interlocking screws, Hofer FixPlug with 3-mm titanium interlocking screw, Hofer Plug with 3.5-mm titanium interlocking screw, and Hofer Plug with 3-mm titanium interlocking screw) in comparison with the commonly used Ender stainless steel nails (locked with 3.5-mm screw) were experimentally investigated in cadaveric Iamb tibiae, regarding their load characteristics and failure modes in the case of heavy loading. The specimens were subjected to sequential axial cyclic loading of 5000cycles with stepwise increase of the load amplitude until failure. Migration of locking screws and internal damage of bone tissue was quantified by micro-computed tomography (CT) imaging. Ender nails failed on average at a peak load of 800 N, TwinPlugs at 1367 N, FixPlugs at 1222 N, Plugs 3.5mm at 1225 N and Plugs 3.0mm at 971 N. TwinPlugs, FixPlugs, and Plugs 3.5mm failed in a slow manner over several hundred loading cycles, whereas Ender nails and Plugs 3.0mm exhibited abrupt failure without any prior indication. Our results confirm that axial stability of FIN can be further improved by screw-fixed plugs by simultaneously avoiding shortcomings of an eye-locked system, which the Ender nails are. Considering biomechanical results, plug interlocking systems with 3.5-mm screws should be favored over conventional Ender nails and plugs with 3-mm screws.

  • 出版日期2016-11-1

全文