Automated thermochemolysis reactor for detection of Bacillus anthracis endospores by gas chromatography-mass spectrometry

作者:Li Dan; Rands Anthony D; Losee Scott C; Holt Brian C; Williams John R; Lammert Stephen A; Robison Richard A; Tolley H Dennis; Lee Milton L*
来源:Analytica Chimica Acta, 2013, 775: 67-74.
DOI:10.1016/j.aca.2013.03.011

摘要

An automated sample preparation system was developed and tested for the rapid detection of Bacillus anthracis endospores by gas chromatography-mass spectrometry (GC MS) for eventual use in the field. This reactor is capable of automatically processing suspected bio-threat agents to release and derivatize unique chemical biomarkers by thermochemolysis (TCM). The system automatically controls the movement of sample vials from one position to another, crimping of septum caps onto the vials, precise delivery of reagents, and TCM reaction times and temperatures. The specific operations of introduction of sample vials, solid phase microextraction (SPME) sampling, injection into the GC MS system, and ejection of used vials from the system were performed manually in this study, although they can be integrated into the automated system. Manual SPME sampling is performed by following visual and audible signal prompts for inserting the fiber into and retracting it from the sampling port. A rotating carousel design allows for simultaneous sample collection, reaction, biomarker extraction and analysis of sequential samples. Dipicolinic acid methyl ester (DPAME), 3-methyl-2-butenoic acid methyl ester (a fragment of anthrose) and two methylated sugars were used to compare the performance of the autoreactor with manual TCM. Statistical algorithms were used to construct reliable bacterial endospore signatures, and 24 out of 25 (96%) endospore-forming Bacillus species were correctly identified in a statistically designed test.

  • 出版日期2013-5-2

全文