摘要

To compare cardiac left ventricular (LV) parameters in simultaneously acquired hybrid fluorine-18-fluorodeoxyglucose ([18F] FDG) positron emission tomography/magnetic resonance imaging (PET/MRI) in patients with residual tracer activity of upstream PET/CT. Twenty-nine patients (23 men, age 58 +/- 17 years) underwent cardiac PET/MRI either directly after a non-cardiac PET/CT with homogenous cardiac [18F] FDG uptake (n=20) or for viability assessment (n=9). Gated cardiac [18F] FDG PET and cine MR sequences were acquired simultaneously and evaluated blinded to the cross-imaging results. Image quality (IQ), end-diastolic (LVEDV), end-systolic volume (LVESV), ejection fraction (LVEF) and myocardial mass (LVMM) were measured. Pearson correlation and intraclass correlation coefficient (ICC), regression and a Bland-Altman analysis were assessed. Except LVMM, volumetric and functional LV parameters demonstrated high correlations (LVESV: r=0.97, LVEDV: r=0.95, LVEF: r=0.91, LVMM: r=0.87, each p < 0.05), but wide limits of agreement (LOA) for LVEDV (-25.3-82.5ml); LVESV (-33.1-72.7ml); LVEF (-18.9-14.8%) and LVMM (-78.2-43.2g). Intra- and interobserver reliability were very high (ICCae<yen>0.95) for all parameters, except for MR-LVEF (ICC=0.87). PET-IQ (0-3) was high (mean: 2.2 +/- 0.9) with significant influence on LVMM calculations only. In simultaneously acquired cardiac PET/MRI data, LVEDV, LVESV and LVEF show good agreement. However, the agreement seems to be limited if cardiac PET/MRI follows PET/CT and only the residual activity is used.

  • 出版日期2017-12